"Haemodynamic effects of intra-aortic balloon pumps stratified by basel" by George Kalapurakal, Vinh Q. Chau et al.
 

Haemodynamic effects of intra-aortic balloon pumps stratified by baseline pulmonary artery pulsatility index

Affiliations

Advocate Christ Medical Center

Abstract

Aims: Intra-aortic balloon pump (IABP) devices are commonly used in patients with heart failure related cardiogenic shock (HF-CS), including those with out-of-proportion right ventricular (RV) dysfunction. Pulmonary artery pulsatility index (PAPi) is a haemodynamic surrogate for RV performance. We aimed to assess short-term haemodynamic changes in patients with HF-CS following IABP support stratified by baseline PAPi.

Methods and results: This is a single-centre study of 67 consecutive patients with HF-CS who underwent IABP placement between 2020 and 2022. The primary aim was haemodynamic changes of specific variables on pulmonary artery catheter monitoring over 72 h following IABP placement. Secondary aims were clinically significant changes in diuretic regimens, changes in inotropes or vasopressors at 72 h following IABP, along with clinical outcomes. Prior to IABP placement, 57% of the total cohort (median age 59 years [48, 69], 31% female) had Society of Cardiovascular Angiography and Interventions Stage C HF-CS. Thirty-eight (56%) patients had a PAPi

Conclusions: IABP devices in those with HF-CS and low or abnormal PAPi may provide modest short-term haemodynamic benefits without significant improvement in CI, along with greater need for adjustment in medical therapeutics to achieve haemodynamic optimization.

Type

Article

PubMed ID

39294848


 

Plum Print visual indicator of research metrics
PlumX Metrics
  • Usage
    • Abstract Views: 5
  • Captures
    • Readers: 3
  • Mentions
    • News Mentions: 1
see details

Share

COinS