Effect of mechanically expanded vs self-expanding transcatheter aortic valve replacement on mortality and major adverse clinical events in high-risk patients with aortic stenosis: The REPRISE III randomized clinical trial

Affiliations

Aurora St Luke's Medical Center

Abstract

Importance: Transcatheter aortic valve replacement (TAVR) is established for selected patients with severe aortic stenosis. However, limitations such as suboptimal deployment, conduction disturbances, and paravalvular leak occur.

Objective: To evaluate if a mechanically expanded valve (MEV) is noninferior to an approved self-expanding valve (SEV) in high-risk patients with aortic stenosis undergoing TAVR.

Design, Setting, and Participants: The REPRISE III trial was conducted in 912 patients with high or extreme risk and severe, symptomatic aortic stenosis at 55 centers in North America, Europe, and Australia between September 22, 2014, and December 24, 2015, with final follow-up on March 8, 2017.

Interventions: Participants were randomized in a 2:1 ratio to receive either an MEV (n = 607) or an SEV (n = 305).

Main Outcomes and Measures: The primary safety end point was the 30-day composite of all-cause mortality, stroke, life-threatening or major bleeding, stage 2/3 acute kidney injury, and major vascular complications tested for noninferiority (margin, 10.5%). The primary effectiveness end point was the 1-year composite of all-cause mortality, disabling stroke, and moderate or greater paravalvular leak tested for noninferiority (margin, 9.5%). If noninferiority criteria were met, the secondary end point of 1-year moderate or greater paravalvular leak was tested for superiority in the full analysis data set.

Results: Among 912 randomized patients (mean age, 82.8 [SD, 7.3] years; 463 [51%] women; predicted risk of mortality, 6.8%), 874 (96%) were evaluable at 1 year. The primary safety composite end point at 30 days occurred in 20.3% of MEV patients and 17.2% of SEV patients (difference, 3.1%; Farrington-Manning 97.5% CI, -∞ to 8.3%; P = .003 for noninferiority). At 1 year, the primary effectiveness composite end point occurred in 15.4% with the MEV and 25.5% with the SEV (difference, -10.1%; Farrington-Manning 97.5% CI, -∞ to -4.4%; P

Conclusions and Relevance: Among high-risk patients with aortic stenosis, use of the MEV compared with the SEV did not result in inferior outcomes for the primary safety end point or the primary effectiveness end point. These findings suggest that the MEV may be a useful addition for TAVR in high-risk patients.

Trial Registration: ClinicalTrials.gov Identifier: NCT02202434.

Document Type

Article

PubMed ID

29297076

Link to Full Text

 

Share

COinS