Heparin-bonded, expanded polytetrafluoroethylene-lined stent graft in the treatment of femoropopliteal artery disease: 1-year results of the VIPER (Viabahn Endoprosthesis with Heparin Bioactive Surface in the Treatment of Superficial Femoral Artery Obstructive Disease) trial

Affiliations

St. Luke’s Medical Center, Aurora Cardiovascular Services

Abstract

PURPOSE: To evaluate the performance of a heparin-bonded, expanded polytetrafluoroethylene (ePTFE)-lined nitinol endoprosthesis in the treatment of long-segment occlusive disease of the femoropopliteal artery (FPA) and to identify factors associated with loss of patency.

MATERIALS AND METHODS: In a single-arm, prospective, 11-center study (VIPER [Gore Viabahn Endoprosthesis with Heparin Bioactive Surface in the Treatment of Superficial Femoral Artery Obstructive Disease] trial), 119 limbs (113 patients; 69 men; mean age, 67 y), including 88 with Rutherford category 3-5 disease and 72 with Inter-Society Consensus for the Management of Peripheral Arterial Disease (TASC II) C or D lesions of the FPA, underwent stent graft implantation. The mean lesion length was 19 cm; 56% of lesions were occlusions. Follow-up evaluations included color duplex ultrasonography in all patients, with patency defined as a peak systolic velocity ratio< 2.5.

RESULTS: At 12 months, Rutherford category and ankle-brachial index (ABI) were significantly improved (mean category improvement, 2.4; ABI increased from 0.6±0.2 to 0.9±0.19; P20% was 70% (P = .047). Primary patency was not significantly affected by device diameter (5 vs 6 vs 7 mm) or lesion length (≤20 cm vs>20 cm). The 30-day major adverse event rate was 0.8%.

CONCLUSIONS: The heparin-bonded, ePTFE/nitinol stent graft provided clinical improvement and a primary patency rate of 73% at 1 year in the treatment of long-segment FPA disease. Careful sizing of the device relative to vessel landing zones is essential for achieving optimal outcomes.

Document Type

Article

PubMed ID

23369553

Link to Full Text

 

Share

COinS