Fenestrated Screws Augmented with PMMA Increase the Pull-out Strength of Sacral Pedicle Screws

Affiliations

Aurora Sheboygan Medical Center

Abstract

STUDY DESIGN:: Biomechanics.

OBJECTIVE:: The objective of this study was to compare the fixation in osteoporotic specimens of a bicortical sacral pedicle screw to the following constructs: a unicortical solid screw augmented with PMMA (TRACT), a unicortical fenestrated screw augmented with PMMA (FEN), and a bicortical screw/sacral alar screw combination (PED/ALA).

SUMMARY OF BACKGROUND DATA:: Rigid posterior sacral fixation of osteoporotic spines is problematic due to poor bone quality. Options to improve fixation include bicortical or tricortical screw trajectory, addition of S2 alar screws, or PMMA augmentation.

METHODS:: 11 osteoporotic cadaveric sacra were potted in two-part filler compound. As a baseline fixation method, a 6.5 mm bicortical pedicle screw was placed into one pedicle of each sacra. The contralateral side was randomly assigned to receive a unicortical fenestrated 6.5 mm pedicle screw augmented with PMMA, a unicortical solid 6.5 mm pedicle screw augmented with PMMA, or a bicortical pedicle screw/alar screw construct. Pedicle screws were failed in cantilever pullout using a materials testing machine. Data were normalized into a test index by taking the ratio of the test screw pullout strength to the intra-specimen bicortical control screw pullout strength.

RESULTS:: The fixation strengths as expressed by the test index of the TRACT, FEN, and PED/ALA groups were 101%, 157%, and 167%, respectively. While there was no statistical difference detected between the FEN and PED/ALA groups, the TRACT group had a trend towards lower pullout strength than the FEN (P=0.06) or PED/ALA (0.06) groups. Although underpowered, this study did not detect biomechanical inferiority for any of the three test configurations when compared to standard bicortical S1 screws.

CONCLUSION:: The data indicate that the fixation strength of a fenestrated, PMMA-augmented construct or a combined S1 pedicle/S2 alar screw construct may be better than either standard bicortical or unicortical PMMA screw-tract augmented screws.

Document Type

Article

PubMed ID

28323708

Link to Full Text

 

Share

COinS